Nebulizer Index Sample Prep Supplies Contact Us Home




NEBULIZER SPECIFICATIONS

All Burgener Nebulizers use Teflon for the sample passage and for the gas orifice.
Most Burgener Nebulizers are 100% Teflon except for the removable gas fittings, but the PEEK Mira Mist and the Ari Mist use PEEK for the outside body, and Teflon for the inner capillaries.
All Burgener Nebulizers use the patented Parallel Path method and the patented Enhanced Parallel Path Method of creating a mist, including the Burgener Legere Nebulizer which uses the Enhanced Method in a Bavington V Groove shape. None of our nebulizers have suction, so liquids must be pumped to the nebulizers.

Standard Nebulizers

Model Body
Material
psi
Operating
Pressure
at .8 l/min
Sample Flow
microliters/min
Sample
Capillary I.D.
microns

Nose Length

Nose Diameter

Overall Length

Standard Gas Line

Pluggable
Appropriate
Chamber
Comments
PEEK
Mira Mist
PEEK 35-45 200 - 2500 425 34 mm 6 mm 67.5 mm 2 mm possible
not likely
All Medium Salts, Particles, excellent %RSD (< .7%), very good sensitivities and excellent long term stability.
Not appropriate for concentrated acids, some organics.
Teflon
Mira Mist
PTFE 35-45 200 - 2500 425 34 mm 6 mm 67.5 mm 2 mm possible
not likely
All Medium Salts, Particles, excellent %RSD (< .7%), very good sensitivities and excellent stability.
Excellent for all acids, all organics.
Micro Flow for alcohol, organics: 50 ul/min +
Graphite
Mira Mist HS
PTFE & Graphite 35-45 200 - 2500 425 34 mm 6 mm 67.5 mm 2 mm possible
not likely
All Medium Salts, Particles, excellent %RSD (< .7%), very good sensitivities and excellent stability. Drip Resistant
Excellent for all acids, all organics.
Micro Flow for alcohol, organics: 50 ul/min +
Ari Mist PEEK 35-45 50 - 1000 75 34 mm 6 mm 67.5 mm 2 mm easy All Medium Salts, NO Particles, excellent %RSD (< .7%), very good sensitivities and excellent stability.
Excellent for all acids, all organics.
Micro Flow for alcohol, organics: 10 ul/min +
TJA Trace PTFE 35-45 800 - 2500 425-800 28 mm 6 mm 58.4 mm 1/4" Tygon with Luer Lock No All High Salts, Particles, good %RSD (< 1%), good sensitivities.
Can not plug nebulizer - inner sample passage larger than capillary used to carry sample from pump to nebulizer.
Only available from TJA / Thermal Element
T2002 PTFE 35-45 800 - 2500 425-800 28 mm 6 mm 58.4 mm 1/8", 1/4",
4 mm
available.
No All High Salts, Particles, good %RSD (< 1%), good sensitivities.
Can not plug nebulizer - inner sample passage larger than capillary used to carry sample from pump to nebulizer.
BTS 50 PTFE 35-45 800 - 2500 425-800 19 mm 6 mm 58.4 mm 1/8", 1/4",
4 mm
available.
No Limited High Salts, Particles, good %RSD (< 1%), good sensitivities
Designed for ARL and older chambers. Works with Scott Double pass chambers. Does not fit cyclonic chambers




Special Purpose Nebulizers

Model Body
Material
psi
Operating
Pressure
at .8 l/min
Sample Flow
microliters/min
Sample
Capillary I.D.
microns

Nose Length

Nose Diameter

Overall Length

Standard Gas Line

Pluggable
Appropriate
Chamber
Comments
Mira Mist - CE PEEK 75-90 2 - 2500 425 34 mm 6 mm 67.5 mm 2 mm possible
not likely
All
or
mini
chamber
Designed for CE / ICP / MS. Allows CE capillary to extend to tip. Also runs as excellent nebulizer without CE capillary.
Ari Mist HP PEEK 65-95 5 - 1000 75 34 mm 6 mm 67.5 mm 2 mm easy All
or
mini
chamber
Medium Salts, NO Particles, good %RSD (< 1%), Excelent sensitivities.
Works with Burgener Mini Chamber - simpler than direct injection with same advantages.
T2002-HP PTFE 75-100 800 - 2500 425-800 28 mm 6 mm 58.4 mm 1/8", 1/4",
4 mm
available.
No All Same as T2002, but with smaller gas orifice. Produces 10% - 30% increase in intensities depending on solutions.
Legere PTFE 80-100 500 - 2500 425-800 18 mm 23.4 mm 31.2 mm 1/8" No Limited High Salts, Particles, excellent %RSD (< .5%), excellent sensitivities
Can run much lower flow rates for organic solvents and alcohols.
Designed for Seiko, ARL and large chambers. Does not fit cyclonic chambers
BTS 100 PTFE 75-100 800 - 2500 425-800 19 mm 6 mm 58.4 mm 1/8", 1/4",
4 mm
available.
No Limited Same as BTS 50 but with smaller gas orifice. Produces 10% - 30% increase in intensities depending on solutions.
High Salts, Particles, better %RSD (< .7%), better sensitivities
Designed for ARL and older chambers. Works with Scott Double pass chambers. Does not fit cyclonic chambers
BTS 30 PTFE 35-45 800 - 2500 425-800 19 mm 6 mm 58.4 mm 1/8", 1/4",
4 mm
available.
No Limited Same as BTS 50, but with larger gas orifice for torches requiring 1.5 l/min flows. High flow torches are not commonly in usage, so this is rarely requested.
LB 30 PTFE Requires ~ 4 L/min
at 35-45 psi
2,000 - 50,000 1000 12 mm 1/4" pipe thread 60 mm 1/8" No Industrial High Sample Flow, excellent mist.
Runs 2 to 50 ml/min, with optimum from 5 to 35 ml/min.
Designed for industrial applications, not for ICP analysis.


Teflon is a registered trademark of E.I. DuPont De Nemours Company and in all cases in this web page refers to PTFE (PolyTetraFluoroEthylene)